4. OPERATING INSTRUCTIONS
After the 553 has been connected into a system accord
ing to the installation information in Section 3, the
operating mode can be selected and the discriminator
thresholds can be adjusted as required for each appli
cation.
Figure 4.1 illustrates the timing relationships that will be
effective in the 553 for each of three possible input pulse
amplitudes. The first two pulses exceed the lower level
threshold without also exceeding the upper level, and the
third pulse exceeds both threshold levels. Superimposed
in each of these input pulses in Fig. 4.1 is an internally-
generated signal that rises to 50% of the input peak
amplitude and is stretched at that level. When the input
pulse decays through the 50% level, the CF discriminator
resets and, when delayed by a fixed interval, triggers the
SCA outputs if the analysis criteria have been satisfied.
The delay interval can be adjusted within the range of 0.1
to 1.1 (us using a front panel control. If the pulses of Fig.
4.1 were furnished to a 553 that was set for its Normal or
Window (differential) mode, time-significant SCA output
pulses would be generated for each of the first two input
pulses but none would be generated for the third be
cause its amplitude was greater than the upper level
threshold. If the pulses were furnished to a 553 that was
set for its Integral mode, a time-significant output pulse
pair would be generated for each of the three input
pulses. In either mode of operation, a ULD output will
result when the corresponding discriminator threshold is
exceeded by the input pulse amplitude.
The adjusted lower level threshold is furnished either
from the front panel Lower Level control or from an
external source through the rear panel LL Ref connector;
the alternate selection is made with the rear panel LL Ref
toggle switch - either Int for the front panel control or
Ext for the rear panel 8NC. In either case, the range is
from 0 to +10 V, measured from chassis ground, of the
input signal.
The adjusted upper level threshold is determined by the
setting of the front panel Upper Level or Window control
and by the front panel mode selector toggle switch.
When the switch is set at either Int or Nor, the range of
the variable control is from 0 to +10 V, measured from
chassis ground. When the toggle switch is set at Win, the
range of the variable control is 0 to +1 V, measured above
the adjusted lower level threshold.
For optimum time-significant SCA output timing, the
front panel Walk Adj must be trimmed to the combined
effects of the input shaping and its time constants and of
the expected dynamic range of operation. Using a dupli
cate of the minimum and maximum amplitudes for
shaped input pulses, set the control to provide the
minimum walk (variation of time of occurrence vs input
amplitude).
ULD
CF TRIGGER
Delay adjusted in range
of 0.1 to 1.1 jjs
—•
(DIFF) SCA POS OUT
(DIFF) SCA NEC OUT
(INT) SCA POS OUT
rL_
(INT) SCA NEG OUT
Y
Hh
-Ih
Hh
■ih
n
Fig. 4.1. Timing Relationships for Input
and Output Pulses.