Fiber test results
may be stored in the FlexScan internal memory or external USB stick.
A
Saved test results are organized into a Fiber Group sub-folder
Results - Fibers
Save As Screen
F
G
Touch Results
to display File Manager and navigate to the desired Project / Fiber Group folder.
•
F
Project, End 1, End 2, and Cable are user-defined in Text Editor.
•
Fiber number auto-increments after each save, but can be modified in Number Editor as needed.
•
Touch Done
when finished.
G
•
For Bi-directional OTDR reports, results must be obtained and stored from each end of the network:
Use same test settings (range, pulse width, etc.) in both directions; Use same launch and receive cables in both
•
directions; Swap End1, End2 names; Bi-directional averaging and report generation is performed using TRM 3.0.
File Manager
Fiber Groups
A
A name of the saved result consists of several parameters, which are
D
defined in the Save As screen
End 1_End 2_Cable_Fiber#_Wavelength
Fiber Group Name
within a Project folder
B
B
E
.
D
E
Fiber#
17
.
C
Projects
S13 for 1310 nm
S15 for 1550 nm
M85 for 850 nm
M13 for 1300 nm
C