2. Installation

2-1. Connecting the MED Multi-well Connector to the MED64 Head
Amplifier
• Read product manuals for all other components of the MED64-Allegro System prior to installation.
1. Place the MED Multi-well Connector on a stable table and AWAY from any equipment which generates
an electric or magnetic filed (e.g. power supply unit, peristaltic pump, heater, refrigerator, etc.).
2. Connects the 4 output terminals of the MED Multi-well Connector to the INPUT terminals of the MED64
Head Amplifier with the MED Connector Cables. Make sure that the terminals with same numbers
are connected. (e.g. 1-16CH to INPUT 1-16CH).
3. Place all cables AWAY from any equipment that generates and electric or magnetic filed.
The MED Connector Cable is very sensitive to noise and vibration. Make sure that the cable
stays away from any equipment that generates electric or magnetic filed, and sits on a stable
table where it cannot be disturbed (i.e. bumped into or displaced during recording).
4. Connects the Ground terminal to the SIGNAL GND in the MED64 Head Amplifier (MED-A64HE1S) with
the green grounding wire to properly ground the Connector.
CAUTION:
Make sure to ground the MED Multi-well Connector to the Head Amplifier. Otherwise considerable exog-
enous noise could be introduced.
Figure 4. Connecting the MED Connector cables (20 pins) to the MED Multi-well Connector.
1-16ch
17-32ch
Figure 5. Connection of the MED Multi-well Connector to the MED64 Head Amplifier (left) and Ground terminal on the
MED Multi-well Connector connected with the grounding wire (right).
33-48ch
49-64ch
Product Manual - MED Multi-well Connector (MED-C11)
2. Installation