2. Installation
2-1. Connecting the MED Multi-well Connector to the MED64 Head
Amplifier
• Read product manuals for all other components of the MED64-Allegro System prior to installation.
1. Place the MED Multi-well Connector on a stable table and AWAY from any equipment which generates
an electric or magnetic filed (e.g. power supply unit, peristaltic pump, heater, refrigerator, etc.).
2. Connects the 4 output terminals of the MED Multi-well Connector to the INPUT terminals of the MED64
Head Amplifier with the MED Connector Cables. Make sure that the terminals with same numbers
are connected. (e.g. 1-16CH to INPUT 1-16CH).
3. Place all cables AWAY from any equipment that generates and electric or magnetic filed.
•
The MED Connector Cable is very sensitive to noise and vibration. Make sure that the cable
stays away from any equipment that generates electric or magnetic filed, and sits on a stable
table where it cannot be disturbed (i.e. bumped into or displaced during recording).
4. Connects the Ground terminal to the SIGNAL GND in the MED64 Head Amplifier (MED-A64HE1S) with
the green grounding wire to properly ground the Connector.
CAUTION:
Make sure to ground the MED Multi-well Connector to the Head Amplifier. Otherwise considerable exog-
enous noise could be introduced.
Figure 4. Connecting the MED Connector cables (20 pins) to the MED Multi-well Connector.
1-16ch
17-32ch
Figure 5. Connection of the MED Multi-well Connector to the MED64 Head Amplifier (left) and Ground terminal on the
MED Multi-well Connector connected with the grounding wire (right).
33-48ch
49-64ch
Product Manual - MED Multi-well Connector (MED-C11)