Problem
f ) One side of image is blurred.
g) Image appears to waver.
h) Field of view becomes only slightly
brighter when the voltage is raised.
2. Electrical System
a) Bulb intermittently lights and goes
out.
b) Bulb burns out almost immediately. Wrong type of bulb is being used.
c) Brightness does not change when
you turn the Light intensity adjust-
ment knob.
d) Voltage indicator LED does not
light or bulb does not light.
3. Coarse/Fine Adjustment
a) Coarse adjustment knob is hard
to turn.
b) Stage drifts down by itself or focus
is lost during observation.
c) Image cannot be focused.
Cause
Recommended immersion oil is not
used.
Dirt/dust on specimen.
Dirt/dust on condenser
Inappropriate object side or cover glass
thickness.
Objective is not correctly engaged in light
path.
Revolving nosepiece is not correctly
mounted.
Stage is not correctly mounted.
Specimen is not correctly mounted on
stage.
Revolving nosepiece is not corrected
mounted.
Objective is not correctly engaged in
light path.
Condenser is not properly centered.
Condenser is not properly centered.
Condenser is lowered too far.
Bulb is nearly burned out.
A connector is not properly connected.
Light intensity preset switch is set to ON. Press switch to OFF.
Bulb is not installed.
Bulb is burned out.
Lamp socket is not connected.
The voltage selector switch is set to the
wrong position.
Tension adjustment ring is tightened ex-
cessively.
You are trying to raise stage without
coarse adjustment knob while pre-focus-
ing lever is kept locked.
Tension adjustment ring is too loose.
When adjusting stage height, you forgot
to reattach upper stopper screw.
Remedy
Use the provided immersion oil.
Clean it.
Replace with glass of recommended
thickness.
Make sure that revolving nosepiece
clicks into place correctly.
Push slide dovetail all the way until it
is stopped.
Re-attach it.
Place specimen correctly on to of
stage and secure it with slide holder.
Push slide dovetail all the way until it
is stopped.
Make sure that revolving nosepiece
clicks into place correctly.
Center the condenser.
Center the condenser.
Adjust the condenser height position.
Replace bulb.
Check all connectors.
Use correct bulb type.
Install bulb.
Replace bulb.
Connect lamp socket correctly.
Set the switch to position matching
your local line voltage (100-120 V or
220-240 V).
Loose ring.
Unlock pre-focusing lever
Tighten ring.
Reattach upper stopper screw.