Olympus BX41 Anleitung Handbuch - Seite 24

Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus BX41 herunter. Olympus BX41 34 Seiten. System microscope
Auch für Olympus BX41: Anleitung Handbuch (36 seiten)

Olympus BX41 Anleitung Handbuch

SPECIFICATIONS

Item
1. Optical system
2. Illumination
3. Focusing
4. Revolving nosepiece
5. Observation tube
6. Stage
UIS (Universal Infinity System) optical system
Built-in transmitted Koehler illumination
6V 30W halogen bulb (pre-centered) 6V30WHAL (PHILIPS 5761)
(Average life time: Approximately 100 hr. when used as directed)
Light intensity voltage range: 2V or less to 5.9 V DC (continuous)
Light intensity preset button (voltage adjustment range: 2V or less to 5.9 V DC)
Rated voltage: 100-120/220-240 V
Power consumption: 45 W
Stage height movement by roller guide (rack & pinion)
Stroke per rotation: 0.1 mm (fine), 17.8 mm (coarse)
Full stroke range: 25 mm
Upper limit stopper
Tension adjustment on coarse focus adjustment knob.
Type
5-position revolving nosepiece
Installable
modules
Type
U-BI30-2
Widefield binocu-
lar tube
Field No.
Tube inclination
30°
I n t e r p u p i l l a ry
distance adjust-
ment
Light path selec-
tor
Type
Common-axis knobs on bottom right. Common-axis knobs on bottom left.
Size
Movement
X- and Y-axis knob with adjustable tension.
mechanism
Movement range: 52 mm in vertical (Y) direction, 76 mm in horizontal (X)
direction.
Specimen hold-
ers (single slide
holder)
Specimen hold-
ers (double slide
holder)
Specification
, 0.8/0.4 A, 50/60 Hz
U-5RE-2
Universal 6-position revolving nose-
piece
Transmitted DIC slider,
None
transmitted light analyzer
U-TBI3
Widefield tilting
Widefield trinocu-
binocular
lar tube
22
5°-35°
continuous
50 mm to 76 mm
None
U-SVRB-4
Rectangular ceramic-coated, wire-driven stage
156 mm (D) x 191 mm (W)
U-HLS4
U-HLST4
U-HLD4
U-HLDT4
U-D6RE
U-TR30-2
U-SWTR-2
Super widefield
trinocular
26.5
30°
3 steps: @ BI 100%
² BI 20%, Photo 80%
³ Photo 100%
U-SVLB-4
U-HRS4
U-HRST4
U-HRD4
U-HRDT4
BX41
22