Olympus CX23 Anleitung Handbuch - Seite 15
Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus CX23 herunter. Olympus CX23 32 Seiten. Biological microscope
6
Adjusting the diopter
a
0
10
20
30
40
50
60
70
} The diopter adjustment is to compensate for the difference of
eyesights between left and right eyes of the observer.
When the eyepiece is not equipped with the eyepiece micrometer
Rotate the diopter adjustment ring a at the upper area of the eyepiece
1
to adjust the reticle to [-1] (both sides).
2
Adjust the interpupillary distance of the eyepiece so that the observation
can be performed with both eyes.
3
Place the specimen.
4
Engage the 10x objective lens into the light path and rotate the coarse/
fine adjustment knob to bring the specimen into focus.
5
Change the objective lens to 40x objective lens, and rotate the coarse/
fine adjustment knob to bring the specimen into focus.
6
Change the objective lens to 10x objective lens. While looking into the
right eyepiece with your right eye, rotate the diopter adjustment ring
to bring the specimen into focus. In the same manner, while looking
into the left eyepiece with your left eye, rotate the diopter adjustment
ring to bring the specimen into focus.
7
Change the objective lens to 40x objective lens again, and rotate the
coarse/fine adjustment knob to bring the specimen into focus.
8
Change the objective lens to 10x objective lens, and confirm that the
right and left eyepieces are focused on the specimen.
9
If the specimen is not focused, bring the specimen into focus according
to the procedure in No.6 and repeat No.7 to 9.
When the eyepiece is equipped with the eyepiece micrometer
1
While looking into the eyepiece which is equipped with the eyepiece
micrometer, rotate the diopter adjustment ring a at the upper area
of the eyepiece to adjust so that the scales and lines of the eyepiece
micrometer in the view field can be viewed clearly.
2
Place the specimen.
3
Engage the 10x objective lens into the light path and rotate the coarse/
fine adjustment knob to bring the specimen into focus.
4
Change the objective lens to 40x objective lens, and rotate the coarse/
80
90 100
fine adjustment knob to bring the specimen into focus.
5
Change the objective lens to 10x objective lens. Rotate the diopter
adjustment ring of the eyepiece which is not equipped with the
eyepiece micrometer to bring the specimen into focus. (Leave the
eyepiece equipped with the eyepiece micrometer as is.)
6
Change the objective lens to 40x objective lens again, and rotate the
coarse/fine adjustment knob to bring the specimen into focus.
7
Change the objective lens to 10x objective lens, and confirm that the
right and left eyepieces are focused on the specimen.
8
If the specimen is not focused, bring the specimen into focus according
to the procedure in No.5 and repeat No.6 to 8.
CX23
11