Problem
9. The stage descends by its own
weight or the focus is lost due to
slippage of the coarse adjustment
knob.
10. The stage can not be lowered
enough.
11. Fields of view of two eyes do not
coincide.
12. Objective hits the specimen
slide when it is switched from
lower magnification to a higher
magnification.
13. The LED illumination does not
light.
14.When changing the magnification,
it is defocused significantly.
Cause
Tension adjustment ring is set too
loose.
The condenser is too much lowered.
The interpupillary distance is not
adjusted correctly.
The difference in diopter of two eyes
are not compensated.
Different eyepieces are used between
the left side and right side.
The specimen slide is set upside
down.
The cover glass is too thick.
The AC adapter or power cord is
unplugged.
The diopter is not matched.
Remedy
Tighten the tension adjustment ring to
get proper tension.
Raise it.
Adjust it correctly.
Compensate it correctly.
Use same eyepiece on the left side
and right side.
Set the specimen slide correctly with
the cover glass up.
Use a cover glass with thickness of
0.17 mm.
Plug it securely.
Adjust the diopter of the eyepiece
correctly.
CX23
Refer-
ence
page
9
10
10
11
22
8
8
20
11
17