Olympus SZX-RFL3 Anleitung Handbuch - Seite 16

Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Laborausrüstung Olympus SZX-RFL3 herunter. Olympus SZX-RFL3 22 Seiten. Fluorescence attachment

ƒ
Fig. 17
A
B
C
D
4. Project the arc image on the stage plate by turning the collector lens
focusing knob ƒ. (A)
...
If the arc image is not protected, adjust the burner centering knobs ....
5. Ring the arc image on the center of the upper (or lower) half of the field by
turning the burner centering knobs .... (B)
6. Focus on the mirror arc image by adjusting the mirror focus screw († in
Fig. 7) on the rear of the lamp housing using the Allen screwdriver. (C)
7. Overlay the arc image with the mirror arc image by turning the burner
centering knobs .... (D)
}During observation, adjust the collector lens focusing knob ƒ so that the
observed field is uniform.
}Hereafter, the mercury burner centering needs not be adjusted until the
next time the mercury burner is replaced.
SZX–RFL3
14