Olympus SZX-RFL3 Instrukcja obsługi - Strona 16
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt laboratoryjny Olympus SZX-RFL3. Olympus SZX-RFL3 22 stron. Fluorescence attachment
ƒ
Fig. 17
A
B
C
D
4. Project the arc image on the stage plate by turning the collector lens
focusing knob ƒ. (A)
...
If the arc image is not protected, adjust the burner centering knobs ....
5. Ring the arc image on the center of the upper (or lower) half of the field by
turning the burner centering knobs .... (B)
6. Focus on the mirror arc image by adjusting the mirror focus screw († in
Fig. 7) on the rear of the lamp housing using the Allen screwdriver. (C)
7. Overlay the arc image with the mirror arc image by turning the burner
centering knobs .... (D)
}During observation, adjust the collector lens focusing knob ƒ so that the
observed field is uniform.
}Hereafter, the mercury burner centering needs not be adjusted until the
next time the mercury burner is replaced.
SZX–RFL3
14