Dinel CLS-23 Manual - Página 10

Navegue en línea o descargue pdf Manual para Accesorios Dinel CLS-23. Dinel CLS-23 20 páginas. Capacitive level sesors
También para Dinel CLS-23: Manual de instrucciones (18 páginas)

Dinel CLS-23 Manual
BN (1)
WH, BU (3)
BN (1)
Fig. 7: "S" type sensors connection
(2 - wire electronic switch)
WH, BU (3)
BN (1)
BK (4)
BU (3)
BN (1)
BK (4)
BU (3)
Fig. 8: "P" (PNP) type sensors connection
Fig. 9: "R" (NAMUR) type sensors connection
Output (BK)
+U (BN)
Fig. 10: Inside of the connector socket
Electrical connection must be done in de-energized state! For switching supply sources,
it is necessary to check that the input is galvanically separated from the network side
and that they are fitted with a filter suppressing the conforming interference (terminals
+ and – oscillate together towards the ground potential), or the interference is removed
in another manner.
10
+U
0V
+U
0V
+U
0V
+U
0V
0V (BU)
CLS–23_–_–_–S–_
Positive  pole  (+  U)  of  power  supply 
is connected through a load (relay) to 
brown (blue) wire or pin connector No. 
1, negative pole is connected to white 
wire or pin connector No. 3. 
Typ CLS–23_–_–_–P–_
Positive pole (+ U) of power supply is 
connected to brown wire or pin connec-
tor No. 1, negative pole is connected to 
blue wire or pin connector No. 3. Load 
(relay) is connected to black wire or pin 
connector No. 4. 
Typ CLS–23_–_–_–R–_
Brown  wire  or  pin  connector  No.  1  is 
connected  to  positive  pole  (+U)  of  In-
trinsically safe supply unit. Blue wire or 
pin.connector    No.  3    is  connected  to 
negative pole of Intrinsically safe sup-
ply unit.
Legend:
(1...3)
– Terminals number for variants with connector
BN
– Brown
WH
– White
BK
– Black
BU
– Blue
CLS–23