To replace the desiccant follow the procedure below.
•
Turn off and unplug the H300 chamber.
•
Remove the desiccant reservoir by gently spreading the holder clamps and pulling the reservoir
away from the back of the chamber. As you pull the two connector hoses will extend.
•
Pull until the hoses will not extend farther. Do not force the hoses to extend. Do not remove the
hoses from the desiccant reservoir. See Figure 2 below.
Figure 2: Desiccant Reservoir Pulled From H300
•
Unscrew the reservoir plug and remove the spring, dust filter and metal mesh.
•
Remove the used up desiccant. The used desiccant can be regenerated by heating it above 395°F
(200°C) for one to two hours.
•
Fill the reservoir with a fresh desiccant. Do not overfill or underfill.
•
Replace the dust filter, followed by the metal mesh, followed by the spring.