To replace the desiccant follow the procedure below.
Turn off and unplug the H300 chamber.
Remove the desiccant reservoir by gently spreading the holder clamps and pulling the reservoir
away from the back of the chamber. As you pull the two connector hoses will extend.
Pull until the hoses will not extend farther. Do not force the hoses to extend. Do not remove the
hoses from the desiccant reservoir. See Figure 2 below.
Figure 2: Desiccant Reservoir Pulled From H300
Unscrew the reservoir plug and remove the spring, dust filter and metal mesh.
Remove the used up desiccant. The used desiccant can be regenerated by heating it above 395°F
(200°C) for one to two hours.
Fill the reservoir with a fresh desiccant. Do not overfill or underfill.
Replace the dust filter, followed by the metal mesh, followed by the spring.