Folyon H300 사용자 설명서 - 페이지 6

{카테고리_이름} Folyon H300에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Folyon H300 16 페이지. Temperature humidity chamber

To replace the desiccant follow the procedure below.
Turn off and unplug the H300 chamber.
Remove the desiccant reservoir by gently spreading the holder clamps and pulling the reservoir
away from the back of the chamber. As you pull the two connector hoses will extend.
Pull until the hoses will not extend farther. Do not force the hoses to extend. Do not remove the
hoses from the desiccant reservoir. See Figure 2 below.
Figure 2: Desiccant Reservoir Pulled From H300
Unscrew the reservoir plug and remove the spring, dust filter and metal mesh.
Remove the used up desiccant. The used desiccant can be regenerated by heating it above 395°F
(200°C) for one to two hours.
Fill the reservoir with a fresh desiccant. Do not overfill or underfill.
Replace the dust filter, followed by the metal mesh, followed by the spring.
H300 Temperature Humidity Chamber
Page 6