Additel ADT927 Manuel d'entretien et de réparation - Page 7
Parcourez en ligne ou téléchargez le pdf Manuel d'entretien et de réparation pour {nom_de_la_catégorie} Additel ADT927. Additel ADT927 15 pages. Hydraulic pressure test pump
Également pour Additel ADT927 : Manuel de l'utilisateur (2 pages), Manuel de l'utilisateur (2 pages), Manuel d'instructions de maintenance (7 pages)
2.1 Pre-pressurization piston
①
Please refer the "Pump cleaning" part of maintenance section to drain the media. Turn the pre-pressurization handle out to the
middle position.
②
Use a 4 mm socket head wrench to remove the three lock bolts on the bottom of the pump.
And use a 24 mm wrench to unscrew the screw nut, as shown in picture 2.1.1.
2.Screw nut
③
Pull out the pre-pressurization piston from the pump chamber, as shown in picture 2.1.2.
④
Please refer to picture 2.1.2 to exchange the damaged parts of the piston.
3. Pre-pressurization
1
handle
Picture.2.1.1
5. Pre-pressurization
piston
4
Picture.2.1.2
6
7
04