CMP6050 Retrofit Installation Instructions
Required Parts and Materials:
F Control panel assembly
F Retrofit cover panel
F Control panel bezel assembly
F Control driver boards
F Control driver board cables
F New/replacement sensors
F Pilaster and light sensor support bracket
F Retrofit Hardware Package
F Spiral wire protection wrap
Procedure:
1.
F
Record information programmed into the controller being replaced prior to beginning installation,
including:
i.
Saved programs
ii.
Schedule
iii.
Network settings
iv.
Alarm Limits
2.
F
Disable power to the chamber by turning OFF the main breaker supplying power from building
electrical service. If the chamber is equipped with a Uninterruptable Power Supply (UPS), ensure it
is also OFF. Switch all breakers in the control panel to the OFF position.
3.
F
Disconnect all cables and wires running from the control panel to the old controller being removed.
Disconnect at the terminal blocks within the control panel.
CONVIRON
F Zip ties
F Tool kit appropriate for servicing Conviron
chambers
F Anti-static wrist strap
F Schematic #231106: SCHEM-Communicaion
Wiring CMP6050
F Schematic #257349: SCHEM-Main Control,
Retro CMP6050 for 4000/5000, 50/60hz
F CMP6050 Operator's Manual
CMP 6050 RETROFIT INSTALLATION INSTRUCTIONS
May 2013 |
257775R00
1