D. Image Acquisition 
Click off the light source button 
 to turn off 
g
 
the non‐laser light.  
d
 
Click on Dye list 
button and double click 
 
 
 
on the fluorescent dyes you want to use (e.g. 
h
 
FITC and TRITC) from the list.  Click Apply 
b
c
button
 (if Assign Dye Manually button
 is 
checked, uncheck before pressing Apply 
button).  It will activate the laser lines and set 
f
detection channels up according to the dyes 
 
you selected.  
 
Choose Image size (default‐ 512x512) from 
d
Size panel
.   
 
Press 
 button (if not pressed) to 
automatically adjust HV and Offset values 
according to the scan speed change.  Click 
e
 button to scan the sample.  It will 
scan fast at 2 us/pixel and show low‐quality 
(pixelated) image in a Live View window.    
b
 
 
While scanning, focus onto the region of interest with the remote focus controller or by clicking arrowhead 
e
 buttons in the microscope window (large arrowhead move the objective by the extent set by StepSize 
and the small arrowheads moves the objective by the half of the large arrowhead step size). 
Click Stop button
 to stop scanning. 
 
f
Set the Laser output level 
 (set the Multi‐Ar laser 
to 1‐5% and the HeNe laser to 1‐25%, depending 
on your sample condition, but no lower than 1%). 
 
g
Set the scan speed 
to 10~12.5 us/pixel and click 
XY repeat 
button to scan (you can scan 
image faster or slower, depending on your sample).   
 
Detector (CHS) Channel Setting 
: Adjust the 
brightness by HV (and Gain) and background black 
level (Offset) of individual channels.  Keep the HV 
value below 700 (beyond 700, the background 
noise will show up).