Special feature of OLYMPUS Micro cantilever
1. Contact mode Cantilever – This type of micro cantilever is suitable for contact mode AFM
measurement.
The small spring constant of the silicon nitride cantilevers enables you to set the tip loading force small
to probe the sample surface gently.
2. AC mode operation in water – This type of micro cantilever is also used for dynamic mode
AFM measurement in water.
For this purpose, 100 m long cantilever in OMCL-TR400PSA- series is recommended.
3. Variety of cantilevers with different mechanical properties – A suitable cantilever can be
chosen from the different types of the cantilevers on a chip.
Each triangular-cantilever type chip (TR type) has two different cantilevers on each end of the chip.
Each rectangular-cantilever type chip (RC type) has four different cantilevers on the chip.
4. Sharpened pyramidal tip
The tip shape about 0.3 micron down from the top is oxide-sharpened so that the tip angle parallel to
the cantilever axis is smaller than 70 deg. and that the tip radius is smaller than 20 nm.
The tip material (silicon nitride) is less prone to wear and tear than a single crystal silicon so that
the tips ensure more stable measurements.
5. Reflex gold coating – Gold is coated on the back side of the cantilevers for optical detection.
Dual side gold coating types are also available, OMCL-TR400PB-, OMCL-TR800PB- and
OMCL-RC800PB-. These make functionalizing the cantilever easier (e.g., using thiol chemistry)
See the specification sheet of OLYMPUS Micro cantilevers at the last page of this manual.
Table of contents
1
Preparation
2
Open the case and take out the cantilever strip from the case
3
Preparation before disconnecting the cantilever chip
4
Disconnecting the cantilever chip
5
Attaching the cantilever chip to your instrument
6
Returning the cantilever strips to the plastic case
7
Storing the cantilever strips in the case for storage
8
Storing the disconnected cantilever chips in the case for storage
9
Information
10
Specification
page
4e
4e
6e
6e
8e
8e
9e
10e
10e
1s
- 3e -
1
Preparation
1) Please prepare the followings before using OLYMPUS cantilevers.
2) To gain a better understanding of how cantilevers and chips are connected, cantilevers should be inspected
under the microscope.
1) Work environment : Clean bench
(Use of an electrical charge neutralizer of ionizer is recommended.)
2) For hazard avoidance : Protective eye glasses, Thin gloves
3) For cantilever treatment and disconnecting procedure
: Microscope slide glass, Tweezers, Scalpel or razor blade, Scotch drafting tape
(Use of anti-electrostatic discharge mat and a wrist band is recommended.)
4) For cantilever storage : Dry box, Case for storage
5) For inspection : Stereo microscope
2
Open the case and take out the cantilever strip from the case
1) It is recommended that the cantilever case be opened in a clean environment like a clean bench in order to avoid
the cantilever being contaminated. Handing under an ionizer is recommended.
2) Avoid wearing clothes like woolen sweaters, fleece, etc that
give off the static electricity when handling the cantilever
cases and chips. Use of an anti-electrostatic mat and wrist
band is preferable.
3) In opening the case, put the plastic case label-side up on a
desk, and turn the case so the arrow mark comes to your
side.
4) Peel off the tape to open the case.
Don't throw away the tape. That will be used again to fix
the closed case after taking out the cantilever strip from
the case.
Caution
 Be sure to grasp the bottom co
case may be opened and the cantilever strips may slip off and break.
5) After peeling off the tape from the case, be sure to grasp the bottom cover like in the drawing left below. Don't
pick up the case by grabbing the cover
Correct
If you pick up the cover like in the
drawing (right), the case may open
and the cantilever strip may slip
off and break
Case
Case
Label
Hold position (Bottom Cover)
Cantilever case (Top view)
ver in taking
the cantilever case, or the
Incorrect
Cover (Case)
Bottom cover (Case)
Cantilever strip
- 4e -
Arrow mark
(indicate opening position)
Lip
Tape
Slip off