Manually Program/Process Quality Control
Consult references for
troubleshooting assistance
UniCel DxC, Customer, Daily Start Up Job Aid
Rev. F (September 2013)
Select Samples from the menu bar
Clear Control IDs
Select Clear F7
Enter the Control ID(s) to be cleared in the
Sample ID(s) field*
Select Enter on the keyboard
Select OK
Select OK to confirm
Manually Program Control
Type the Control ID in the Sample ID field
Select Enter on the keyboard
Optional: Enter rack/position number if a bar code
is not available
Select the chemistries to be run
Select Save F10
Repeat steps to program another level, if necessary
Pour the control material into sample cup(s). Nest the
sample cup(s) in the appropriate bar code labeled
tube(s) or place in assigned rack/position(s)
Place the rack in the autoloader with the rack bar code
facing to the right
Press the <RUN> button on the instrument
When Quality Control is complete,
review for acceptability (defined by laboratory policy)
Note: If a manually programmed control is incomplete,
the system will only try to run the incomplete tests the
next time you run the Control ID. Clear the Control ID
to clear the incomplete QC sample and
reset Auto Generation of Control
No
Quality Control
acceptable?
*Clear multiple IDs
by typing in each
Control ID
separated by a
Yes
Procedure complete
Beckman Coulter Professional Development
comma
Page 13 of 15