Beckman Coulter UniCel DxC 600 Mulai Harian - Halaman 14
Jelajahi secara online atau unduh pdf Mulai Harian untuk Peralatan Medis Beckman Coulter UniCel DxC 600. Beckman Coulter UniCel DxC 600 15 halaman. Synchron clinical system
- 1. Table of Contents
- 2. Clear Manually Assigned Racks
- 3. Check Reagent Status
- 4. Load MC Reagents
- 5. Load CC Reagents
- 6. Unload CC Reagents
- 7. Check Calibration Status
- 8. Calibrate
- 9. Process Quality Control Using Auto Generation of Control
- 10. Manually Program/Process Quality Control
- 11. Sample Processing
- 12. Low Volume Sample Processing
Remove visible blood from the top of
capped tubes using a cotton tipped
swab moistened with DI water
Select Samples
from the menu bar
Enter the Sample ID. Enter Rack
and Position
Select Next F10
UniCel DxC, Customer, Daily Start Up Job Aid
Rev. F (September 2013)
Sample Processing
Start Here
Yes
Validated
CTS primary
*Adequate
sample
Adequate
volume
sample
determined by
Primary Tube
volume*?
Sample
Template
Yes
Is sample
programmed at
the LIS?
Yes
Is there a
No
readable bar
code?
Yes
Place container in the
appropriate rack. Ensure
that bar code label, if
present, is visible through
the rack slot. Verify the
tube is seated properly
Load the prepared racks
on the autoloader with the
rack bar code facing to the
Press the <RUN> button
on the instrument
No
tube?
No
No
right
Beckman Coulter Professional Development
Remove cap
Use the
Low Volume Sample
Processing
flowchart
Select Samples
from the menu bar
Enter the Sample ID.
Enter Rack and Position if
readable bar code is not available
Select chemistries to be run
If necessary, select STAT check
box, Sample Type, Sample
Comment and/or Demog F2
Select Next F10
Page 14 of 15