- 1. Table of Contents
- 2. Clear Manually Assigned Racks
- 3. Check Reagent Status
- 4. Load MC Reagents
- 5. Load CC Reagents
- 6. Unload CC Reagents
- 7. Check Calibration Status
- 8. Calibrate
- 9. Process Quality Control Using Auto Generation of Control
- 10. Manually Program/Process Quality Control
- 11. Sample Processing
- 12. Low Volume Sample Processing
Select Samples
from the menu bar
Enter the Sample ID. Enter Rack
and Position
Select Next F10
UniCel DxC, Customer, Daily Start Up Job Aid
Rev. F (September 2013)
Low Volume Sample Processing
programmed at the
No
readable bar
Transfer the sample into a validated insert cup and
nest the cup in the bar coded primary tube.
Place in a Reserved rack
Transfer the sample into a 0.5 mL sample cup
Note: Nest the 0.5 mL sample cup in a red metal
insert to reduce the dead volume to 20 µL.
Place the nested red metal insert in a Reserved
rack. Bar code label can be attached to the red
Load the prepared rack(s) on the autoloader with
the rack bar code facing to the right
Press the <RUN> button on the instrument
Is sample
No
LIS?
Yes
Is there a
code?
Yes
OR
metal insert
Beckman Coulter Professional Development
Program chemistries
using instructions on
Sample Processing
flowchart