DSI MINI SURGERY MOTOR Panduan Pengguna - Halaman 7

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Peralatan gigi DSI MINI SURGERY MOTOR. DSI MINI SURGERY MOTOR 15 halaman. Micromotor for oral surgery

USER MANUAL
Micromotor for oral surgery
01_26/04/2018
• Connect the irrigation tube with the physiological solution bag and with the handpiece.
• Connect the foot-switch (or the foot-control) to the panel connector.
• Place the pedal in a easy-to-use position.
• Check that main power switch is in 0 position (off).
• Connect the power supply cord to the socket in the back panel of control unit.
• Connect the power supply cord to a grounded socket.
• Switch on the device by the main switch.
USE
 All the adjustments are to be made with the motor stopped.
The display shows the following elements:
SPEED: rotation speed (max 40000 rpm with reduction ratio 1:1);
TORQUE: maximum torque (max 80 Ncm with reduction ratio ≥ 32:1);
PUMP: physiological solution flow rate (max 90 ml/min);
REDUCT: reduction ratio of the current use handpiece;
PROGRAM: the current use program (1, 2, 3......, 10);
Symbol
: led activation;
Symbol
: activation of "reverse" function (rotation way changing);
Symbol
: pump activation.
Pic. 6.
Page 7 of 15