DSI MINI SURGERY MOTOR Gebruikershandleiding - Pagina 7

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} DSI MINI SURGERY MOTOR. DSI MINI SURGERY MOTOR 15 pagina's. Micromotor for oral surgery

USER MANUAL
Micromotor for oral surgery
01_26/04/2018
• Connect the irrigation tube with the physiological solution bag and with the handpiece.
• Connect the foot-switch (or the foot-control) to the panel connector.
• Place the pedal in a easy-to-use position.
• Check that main power switch is in 0 position (off).
• Connect the power supply cord to the socket in the back panel of control unit.
• Connect the power supply cord to a grounded socket.
• Switch on the device by the main switch.
USE
 All the adjustments are to be made with the motor stopped.
The display shows the following elements:
SPEED: rotation speed (max 40000 rpm with reduction ratio 1:1);
TORQUE: maximum torque (max 80 Ncm with reduction ratio ≥ 32:1);
PUMP: physiological solution flow rate (max 90 ml/min);
REDUCT: reduction ratio of the current use handpiece;
PROGRAM: the current use program (1, 2, 3......, 10);
Symbol
: led activation;
Symbol
: activation of "reverse" function (rotation way changing);
Symbol
: pump activation.
Pic. 6.
Page 7 of 15