DSI MINI SURGERY MOTOR 사용자 설명서 - 페이지 7

{카테고리_이름} DSI MINI SURGERY MOTOR에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. DSI MINI SURGERY MOTOR 15 페이지. Micromotor for oral surgery

USER MANUAL
Micromotor for oral surgery
01_26/04/2018
• Connect the irrigation tube with the physiological solution bag and with the handpiece.
• Connect the foot-switch (or the foot-control) to the panel connector.
• Place the pedal in a easy-to-use position.
• Check that main power switch is in 0 position (off).
• Connect the power supply cord to the socket in the back panel of control unit.
• Connect the power supply cord to a grounded socket.
• Switch on the device by the main switch.
USE
 All the adjustments are to be made with the motor stopped.
The display shows the following elements:
SPEED: rotation speed (max 40000 rpm with reduction ratio 1:1);
TORQUE: maximum torque (max 80 Ncm with reduction ratio ≥ 32:1);
PUMP: physiological solution flow rate (max 90 ml/min);
REDUCT: reduction ratio of the current use handpiece;
PROGRAM: the current use program (1, 2, 3......, 10);
Symbol
: led activation;
Symbol
: activation of "reverse" function (rotation way changing);
Symbol
: pump activation.
Pic. 6.
Page 7 of 15