Operating instructions

Environmental conditions

This apparatus is intended for indoor use only.
This apparatus can be operated safely at an altitude up to 2000 m.
The normal operating temperature range is between 4°C and 65°C.
Maximum relative humidity 80 % for temperatures up to 31°C decreasing linearly to 50 % relative
humidity at 40°C.
The apparatus is rated Pollution Degree 2 in accordance with IEC 664. Pollution Degree 2 states
that: "Normally only non-conductive pollution occurs. Occasionally, however, a temporary conductivity
caused by condensation must be expected".

Setting up the horizontal gel tank

Fitting electrode cables
Note the position of the lid on the unit. This shows the correct polarity and the correct orientation of
the cables, black is negative and red positive.
Remove the lid from the unit. Note if the lid is not removed, fitting the cables may result in un-
tightening of the gold plug and damage to the electrode.
Screw the cables into the tapped holes as fully as possible so that there is no gap between the lid
and the leading edge of the cable fitting.
Refit the lid.
Fitting loading guides
These can be fitted to enhance visibility of the wells if desired. They can be fitted to the white vinyl platform
sheet or to the unit itself.
Seat the tray in the unit and note the position of the comb grooves. The samples run black to red but
the trays can be used frontward or backwards so ensure that the comb grooves closest to the black
electrode are marked.
Remove the tray.
Peel the back off of the loading guide and carefully apply the loading guide directly to the gel platform.
The unit is now ready to be used.
Horizontal Gel Electrophoresis units
2