MU-1
• Criteria
The resistance between the external case and grounding line should be 0.2 ohm or less.
• Inspection procedure
Measure the resistance between the grounding terminal in the removable power supply
cord and the screw that is farthest from the inlet.
See the inspection procedures for grounding resistance testing instrument for the
measurement methods.
Safety Check
3. Grounding Resistance
6-2