Date of inspection
Product
Serial number
Electrical Safety Inspection
Item
2-1-1
External leak current
2-1-2
Grounding leak current
2-1-3
Grounding Resistance
ISSUE6
4. Safety Check Sheet
MU-1
Between the external case and grounding line.
Inspected by
Checked by
Criteria
Normal Condition
Single Fault
Normal Condition
Single Fault
6-3
MU-1
Result
100 A or less
500 A or less
500 A or less
1000 A or less
0.2 Ω or less