AKSO Cor Tester Manuale di istruzioni - Pagina 5

Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di prova AKSO Cor Tester. AKSO Cor Tester 8.

9) With the instrument in the VERTICAL position, biefly
press the button, the message Analyzing reference will
appear on the display;
10) Wait until the display displays the Insert sample
message;
Reading – Apparent Color
While Insert sample is on the display:
11) Open the measuring cover and remove the cuvette
from the compartment;
12) With the aid of the syringe, collect al least 15mL of the
sample to be analyzed;
13) With the sample in the syringe, fill in the glass cuvette
(12ml);
14) Close the cuvette with its cover;
15) Clean and dry puter wall of the cuvette using a cleaning
wipe or soft towel paper;
16) Insert the cuvette into the measuring compartment;
17) Measuring alignment: for better accuracy insert the
cuvette always in the same position;
18) Close the measuring cover on the glass cuvette to
avoid interference in the readings;
19) Briefly press the Power-Measuring button and wait.
The display will display the Message Analyzing
Sample, followed by the measurement result in PCU;
20) At the end of the measurement, discard the sample
analyzed and immediately wash the glass cuvette with
the distilled or deionized water.