Olympus Fluoview FV1000 Manuale d'uso - Pagina 31

Sfoglia online o scarica il pdf Manuale d'uso per Microscope Olympus Fluoview FV1000. Olympus Fluoview FV1000 42. Confocal microscope
Anche per Olympus Fluoview FV1000: Panoramica (28 pagine), Istruzioni brevi (7 pagine), Istruzioni brevi (7 pagine), Manuale d'uso (9 pagine), Manuallini (3 pagine)

Olympus Fluoview FV1000 Manuale d'uso

3 Centration of Mercury Burner

3-1 Centering the Mercury Burner

For the reflected light fluorescence observation, refer to the User's Manual for the
Reflected Light Fluorescence System.
Since this system introduces the light of a mercury burner through the light guide, the
burner centering method is slightly different from that described in the User's Manual for
the Reflected Light Fluorescence System. This section is intended to describe the
method specific to this system.
1.
2.
3.
TIP
The arc image stabilizes in 5 to 10 minutes after the mercury burner is turned on.
NOTE
Keep the shutter
target from being heated up.

Centration of Mercury Burner

Turn the shutter
fully toward the bottom to block the light.
When the light guide is disconnected, another built-in shutter is engaged
automatically in the light path to ensure safety.
Remove the light guide
target
.
Turn the shutter
toward the open direction. The arc image of the mercury burner
will be visible on the screen
closed except for centering operation to prevent the centering
/ Centering the Mercury Burner
from the ULH holder and replace with the centering
of the centering target
.
II. PREPARATION For OPERATION
Page
II. 3 - 1