Olympus CIX100 Manuale di configurazione dei sistemi - Pagina 8

Sfoglia online o scarica il pdf Manuale di configurazione dei sistemi per Apparecchiature di laboratorio Olympus CIX100. Olympus CIX100 38. Cleanliness inspector system
Anche per Olympus CIX100: Manuale di installazione (14 pagine), Manuale di istruzioni (46 pagine)

Olympus CIX100 Manuale di configurazione dei sistemi
OLYMPUS Cleanliness Inspector CIX100
2 Safety instructions
2.5

Avoiding device damage

Attention
Device damage through manual operation of the stage
The stage can be damaged by manual operation.
Use the joystick to move the stage or to change its height.
Attention
Device damage through overloading the stage
Overloading the stage can hinder the movement of the stage or damage the
microscope.
For the inspection, only use membrane filters that are mounted in the filter
holder.
Attention
Device damage from oversize or unsuitable samples
Samples that aren't clamped on the filter holder or that exceed the
maximum permitted height can damage the lenses or the stage.
For the analysis, only use membrane filters that are mounted in the filter
holder.
Make sure that the height of the sample doesn't impair the movement of the
objective.
System Setup Guide
8