Olympus CIX100 Manuale di configurazione dei sistemi - Pagina 9

Sfoglia online o scarica il pdf Manuale di configurazione dei sistemi per Apparecchiature di laboratorio Olympus CIX100. Olympus CIX100 38. Cleanliness inspector system
Anche per Olympus CIX100: Manuale di installazione (14 pagine), Manuale di istruzioni (46 pagine)

Olympus CIX100 Manuale di configurazione dei sistemi
2.6

Notes on placement

• The weight of the microscope system is approx. 18 kg. Consider this
• Place the microscope system on a sturdy level table or bench, taking care
• When setting up the microscope system, ensure that there is a minimum
Attention
Device damage through improper lifting of the microscope system.
The microscope system may be damaged if you lift it by an unsuitable
component, the stage for example. Furthermore, the calibrations can
become invalid.
When transporting the microscope system, hold it carefully by the grip on
the arm (see figure 1).
Figure 1
2.7

Notes on using the system with networks

If the computer that the software is installed on is part of a network
environment, note the following points.
• Make sure that the network has appropriate virus protection.
• The installation of operating system service packs, security patches or
• If anti-virus software is run while you are acquiring images, it can lead to
weight when selecting the work surface and method for lifting the
microscope system.
not to block the ventilation slots on the underside of the frame.
distance of 10 cm from the wall and other devices.
third party software updates (including runtime libraries) can cause
software malfunctions.
the loss of individual acquisitions. It can also slow down the acquisition
process or even cause it to be aborted altogether.
9
OLYMPUS Cleanliness Inspector CIX100
2 Safety instructions
System Setup Guide