Step 2: Install the DPPMCA Display and Acquisition Software
Locate the DPPMCA directory and copy and paste it to your local machine. Then run the executable in the
directory. There is no "setup," it is a simple copy and paste to the local computer. Amptek recommends copying the
software to "Program Files\Amptek."
TIP: Right click on the DPPMCA.exe file and select create shortcut. Then drag the shortcut to your desktop for
easy access.
You must have full read/write access to the DPPMCA directory. It cannot run from the CD. In Windows 7 it is
best to right-click the executable and select Run as Admin.
MCA8000D Connection Instructions
1. Verify that the MCA8000D is completely unplugged (both USB and power).
2. Use the USB cable to connect the MCA8000D USB connector to the PC's USB connector. When using the USB
connection the unit is powered through USB, the external +5V AC adapter is not needed.
NOTE: Grounding is critical to obtaining the best performance. We strongly recommend using a single point
ground for the system. To ensure this, use a 3-prong to 2-prong adapter on the power supply of any
notebook computer. This prevents the notebook from introducing another ground into the system.
Quick Use of the MCA8000D
1. Connect the MCA8000D and PC as described in the previous section.
The MCA8000D stores its configuration, so when powering the system initially, the factory configuration is
loaded. If you modify the configuration, the last configuration is loaded.
2. Launch the DPPMCA software by opening the DPPMCA.exe file.
3. Connect to the MCA8000D
a. The connection dialog will appear when DPPMCA is launched.
b. Select which interface you will use (USB, Ethernet, or RS232).
c. If more than one device is available, select the device you want.
d. Click "connect."
4. Configure the MCA8000D
a. The most commonly changed parameters are the number of channels and the low level discriminator (LLD)
threshold.
 To quickly change the LLD click the cursor to the desired channel and press F8. Then press the A key on
the keyboard to clear the data.
b. For all configuration parameters, click
DPP Setup. Once you have changed a parameter, click the "Apply" button to send it to the MCA8000D.
c. Within the DPP Setup dialog box, you can load a default configuration file or you can save and load custom
configurations.
5. Acquire data
a. To start or stop acquisition, click the "start/stop" icon in the toolbar or press the space bar.
b. To clear the data and time, click the "delete data and reset time" button in the toolbar or press "A" on the
keyboard.