- If necessary, press ZERO key to set the display to "0"
- Press QTY/SET key, display will show CH=OFF or CH=ON
- Press the ENTER key to select "ON"
- Press [*] key, display will show a flashing "0" at UNIT WEIGHT, key in the lower limit using
the numeric key pad, press [*] gain, display will show a flashing "0" at TOTAL, key in the upper
limit using the numeric keypad.
-Press QTY/SET key to confirm the setting and return to weighing mode.
- Press the SCALE/SELECT key to select either SCALE 1 or SCALE 3, be sure to place
the additional pieces on the scale that is selected.
***NOTE: When adding items onto the scale, the scale will beep in a particular tone if the
weight (number of items) is less than the lower limit and beep in another tone if it is higher than
the upper limit.

5 Calibration - Using an External Calibration Weight:

Calibration is required when the weighing machine is initially installed or if the scale is moved to a
substantial distance from the original location. 10 minutes warm up of scale is always needed before
calibration.
5.1 Scale 1 Calibration
- Turn on scale
- As the scale is counting down, press the following buttons in order [1], [3], [1], [1]
then press ENTER
- Display will show the following: 'WEIGHT DISPLAY' shows SCALE 1, 'UNIT WEIGHT'
shows CAL-0, 'TOTAL DISPLAY' will show A/D value
- Press UNIT key to select whether you will calibrate in kg or lb
- Press ENTER to set zero point (UNIT WEIGHT will show CAL-0 flashing for a
few seconds and then changes to XXXX to set the calibration weight)
- Press [CE] key and then press the corresponding number buttons based off of the
amount of weight you are using to calibrate (example: if you are using 5lbs to calibrate the
scale you would key in 5lbs).
- Calibration weight will show in the UNIT WEIGHT display.
- Press the ENTER key, the UNIT WEIGHT display will show flashing XX and then
changes to XX.XXX lb. Once this occurs the calibration process is complete.