two adjustment screws with the 5/32 allen wrench.
Step 6 - Installing the cover & left fine focus knob
Remove the two screws from the edges of the drive assembly. Locate the ASI cover plate. Place it in
position over the motor drive assembly and secure it with the screws just removed. Place the left fine
focus knob over the protruding end of the drive shaft and secure it in place with the original screw
using the 2.5 mm allen wrench. Install the 20mm black disk that covers the fine focus knob
attachment screw. This 20mm black disk was removed in step #1.
Connecting the motor drive to the controller and computer
Please refer to the wiring drawings and diagrams given in the manual
Tools Needed: Small slotted screw driver.
1) Locate the drive cable. Connect one end of the drive cable to the pigtail extending from the motor
drive near the microscope and the other end to mating connector on the back of the Microscope
Focus Controller, or MS-2000 unit labeled "Drive". Secure the cable to the drive by tightening the
screws at each end of the connector.
2) Locate the computer interconnect cable. Connect one end to the mating connector on the back of
the controller labeled "Serial In". Connect the other end to the appropriate serial output connector on
your computer. .
3) Plug in the power cord / power supply
This completes the steps necessary to interconnect the ASI motor drive, focus controller and
computer. Please refer to the operation section of this manual before using the unit