- ページ 20

顕微鏡 Olympus BX51のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus BX51 40 ページ。 System
Olympus BX51 にも: 取扱説明書 (17 ページ), 取扱説明書 (34 ページ), 取扱説明書 (37 ページ), 取扱説明書 (11 ページ)

Olympus BX51 取扱説明書
16
@
²
Fig. 25
@
Fig. 26
4

Using the Eyepiece Micrometer Disk

When the WHN10X-H (or WHN10X) eyepieces are used, an eyepiece
micrometer disk can be inserted in one of them. When the eyepiece
does not have a diopter adjustment mechanism, however, it is hard to
focus on the micrometer disk if the operator has poor eyesight. Should
that be the case, adjust the focus with eyeglasses on.
Use 24 mm dia. x 1.5 mm thick micrometer disks.
Following Fig. 25, turn the built-in micrometer-mounting frame ² counter-
clockwise to remove it from the eyepiece and place a micrometer disk.
Screw the micrometer mounting frame back into the eyepiece sleeve.
Selecting the Light Path of the
5
Trinocular Tube
Slide the light path selector knob @ to select the desired light path.
Trinocular tube
Pushed in
U-TR30-2
U-SWTR-3
100% for binocu-
lar eyepieces
U-TR30NIR *
* With the infrared trinocular tube, infrared observation up to 1000 nm is
possible. For details, consult your Olympus representative.
Light path selector knob position
Middle position
20% for binocular
eyepieces, 80% for
TV/photography
50% for binocular
eyepieces, 50% for
TV/photography
(Fig. 25)
(Fig. 26)
Pulled out
100% for TV/
photography