8.
Exit pupil center check
8-1
Set the following jigs as illustrated below to check the exit pupil center.
• Eyepiece------ KN0029: Centering telescope
• Objective ----- MDPlan5X
• Other --------- BKN0003: Standard tube
Make sure that the displacement between the center of the objective's exit pupil diameter and the
cross hairs center of centering telescope is within 15% of the objective's exit pupil diameter.
If the above requirement is not met, the MIRROR • is tilted (failure in installation) or defective.
C. REPAIR PROCEDURE
Mirror surface
C-14
Centering
telescope
(KN0029)
Standard tube
(BKN0003)
MDPlan5X
•