Olympus CK40M FRAME Onarım Kılavuzu - Sayfa 35

Mikroskop Olympus CK40M FRAME için çevrimiçi göz atın veya pdf Onarım Kılavuzu indirin. Olympus CK40M FRAME 38 sayfaları. Compact inverted metallurgical microscope

Olympus CK40M FRAME Onarım Kılavuzu
CK40M
8.

Exit pupil center check

8-1
Set the following jigs as illustrated below to check the exit pupil center.
• Eyepiece------ KN0029: Centering telescope
• Objective ----- MDPlan5X
• Other --------- BKN0003: Standard tube
Make sure that the displacement between the center of the objective's exit pupil diameter and the
cross hairs center of centering telescope is within 15% of the objective's exit pupil diameter.
If the above requirement is not met, the MIRROR • is tilted (failure in installation) or defective.
C. REPAIR PROCEDURE
Mirror surface
C-14
Centering
telescope
(KN0029)
Standard tube
(BKN0003)
MDPlan5X