- ページ 26

産業機器 Siemens SITRANS WFS300のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Siemens SITRANS WFS300 42 ページ。 Sensing heads

Siemens SITRANS WFS300 取扱説明書
Commissioning

6.5 Integrator calibration

6.5
Integrator calibration
Refer to the Integrator operating instructions for integrator calibration instructions.
6.6

Zero calibration

Refer to the Integrator operating instructions for zero calibration instructions.
6.7

Span calibration

After a successful zero calibration, apply the test weight for the span calibration:
1. Attach one end of a string (monofilament fishing line or fine flexible cable) to the test
weight.
2. Route the other end of the string over the calibration pulley.
3. Attach the free end of the string to the string fixing bolt.
26
Sensing plate
Changes < 0.01 V AC
Voltmeter
Test weight
Operating Instructions, 03/2021, A5E32880243-AC
SITRANS WFS300