Siemens SITRANS WFS300 사용 설명서 - 페이지 26

{카테고리_이름} Siemens SITRANS WFS300에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Siemens SITRANS WFS300 42 페이지. Sensing heads

Siemens SITRANS WFS300 사용 설명서
Commissioning

6.5 Integrator calibration

6.5
Integrator calibration
Refer to the Integrator operating instructions for integrator calibration instructions.
6.6

Zero calibration

Refer to the Integrator operating instructions for zero calibration instructions.
6.7

Span calibration

After a successful zero calibration, apply the test weight for the span calibration:
1. Attach one end of a string (monofilament fishing line or fine flexible cable) to the test
weight.
2. Route the other end of the string over the calibration pulley.
3. Attach the free end of the string to the string fixing bolt.
26
Sensing plate
Changes < 0.01 V AC
Voltmeter
Test weight
Operating Instructions, 03/2021, A5E32880243-AC
SITRANS WFS300