Wait 20 minutes after the test subject smokes, eats or drinks
before testing. Failure to wait 20 minutes can give inaccurate BAC
readings and harm the sensor.
Step 1) Place a fitted cover over the mouthpiece each time you test
another individual.
Step 2) If it has been more than 24 hours since the device was last
used, prime the sensor by blowing into the mouthpiece for 4
to 5 seconds before turning on the device.
Step 3) PRESS the red Power On/Off button located on the front of the
unit. A single beep will sound and the device will count down
from 99 as it prepares the sensor for testing.
Note: The device will restart a count down from 10 if the sensor is not
ready within the first countdown from 99.
Step 4) WAIT until the unit displays "bL" and you hear a SINGLE
BEEP.
Note: The device will not shut off until a test is performed OR until 15
seconds of non-use elapses.
Step 5) Take a DEEP BREATH AND BLOW STEADILY (for at least
5 seconds) until a BEEP sounds to signal completion of the
testing.

OPERATION

5