Doran 4300 운영자 매뉴얼 - 페이지 7

{카테고리_이름} Doran 4300에 대한 운영자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Doran 4300 12 페이지.

Basic Checkweighing Reference Guide
Over, Under and Accept Checkweighing Operation:
1) Remove all items from the scale platter.
2) Press ZERO to zero the scale. The weight indication should now be zero.
3) Place an item on scale platter and wait for the motion (MOT) to turn off,
indicating a stable weight.
4) If the item is heavier than the over limit, the OVER indicator will light, If the
item is lighter than the under limit, the UNDER indicator will light. If the weight
is between the limits, the ACCEPT indicator will light.
Digital Entry and Recall of Over and Under Limits:
1) Press and release the OVER button. The OVER and ACCEPT indicators will
light followed by momentary "over" message.
2) Current weight value of saved limit will be displayed.
3) The scale is in the scroll mode. Press either the OVER or UNDER button to
increase or decrease the limit value. Holding a button longer will cause the
count to accelerate. Press the UNITS or PRINT button to cancel the check
limit value entry. The display will read "abort."
4) Once the desired limit is reached, press ZERO to save the limit. The display
will read "donE."
5) To enter the under limit, press and release the UNDER button. The UNDER
and ACCEPT indicators will light followed by momentary "under" message.
Then repeat steps 2 to 4.
NOTE: The model 4300 is configured at the factory for the digital entry of over
and under limits. Other entry configurations are available, refer to your dealer or
Instruction manual for details.
Digital Entry of Over and Under Limits: (optional configuration)
1) Zero the scale.
2) If available, place an item with the desired weight on the scale.
3) Press and release the OVER button. The OVER and ACCEPT indicators will
light followed by momentary "over" message.
4) The scale is in the scroll mode. Press either the OVER or UNDER button to
increase or decrease the limit value. Holding a button longer will cause the
count to accelerate. Press the UNITS or PRINT button to cancel the check
limit value entry. The display will read "abort."
5) Once the desired limit is reached, press ZERO to save the limit. The display
will read "donE."
6) To enter the under limit, press and release the UNDER button. The UNDER
and ACCEPT indicators will light followed by momentary "under" message.
Then repeat steps 4 and 5.
7