Andon KD-322 운영 매뉴얼 - 페이지 14

{카테고리_이름} Andon KD-322에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Andon KD-322 14 페이지. Semi automatic arm cuff blood pressure monitor

DN:KD-322-SMSY01 V4.0
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the KD-322 as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output
power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
14/14
m
80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
Operation Guide
800 MHz to 2,5 GHz
d
2
3 .
P
0.23
0.73
2.3
7.3
23
Template version:1