electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the KD-322 as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output
power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
14/14
m
80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
Operation Guide
800 MHz to 2,5 GHz
d
2
3 .
P
0.23
0.73
2.3
7.3
23