Danatronics ECHO 7DL 빠른 시작 매뉴얼 - 페이지 6

{카테고리_이름} Danatronics ECHO 7DL에 대한 빠른 시작 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Danatronics ECHO 7DL 9 페이지. Thickness gages

Danatronics ECHO 7DL 빠른 시작 매뉴얼
Make any necessary adjustments then F4 for to go to the Measurement screen
Saving and discarding an altered stored setup:
When making a change to any parameter to a default setup, the ECHO will put an asterisk next to
the file indicating something has changed. Your choice is to Save the new setup by pressing F2
and typing in a new name or discarding the changes by pressing F3. The ECHO can save more
than 2,600 custom setups which is ideal for varying applications.
Note: The default setups will not be erased when assigning a new Setup File
4. Calibrating the Gage
Calibrating is the process of adjusting the gage for a specific material and transducer before
testing the material to make sure that all measurements are accurate. You must always calibrate
before measuring material for standard accuracy. The following steps show how to perform
velocity cal, zero cal, velocity and zero cal, and delay-line cal. A test step block of known
thicknesses is required to perform an accurate calibration.
Ver: 2.0
6