Anyload EC200 운영 매뉴얼 - 페이지 14

{카테고리_이름} Anyload EC200에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Anyload EC200 16 페이지.
Anyload EC200에 대해서도 마찬가지입니다: 사용자 설명서 (8 페이지)

Anyload EC200 운영 매뉴얼
-
***Note: Make sure nothing is loaded on the platform before you are asked to place the
calibration weight. Once calibration is complete you will need to reboot the scale and place the
sample test weight on the platter to assure the calibration was completed correctly.
5.2 Scale 2 Calibration
- Turn on scale
- As the scale is counting down, press the following buttons in order [1], [3], [1], [2]
then press ENTER
- Display will show the following: 'WEIGHT DISPLAY' shows SCALE 2, 'UNIT WEIGHT'
shows CAL-0, 'TOTAL DISPLAY' will show A/D value
- Press UNIT key to select whether you will calibrate in kg or lb
- Press ENTER to set zero point (UNIT WEIGHT will show CAL-0 flashing for a
few seconds and then changes to XXXX to set the calibration weight)
- Press [CE] key and then press the corresponding number buttons based off of the
amount of weight you are using to calibrate (example: if you are using 1lb to calibrate the
scale you would key in 1lb).
- Calibration weight will show in the UNIT WEIGHT display.
- Press the ENTER key, the UNIT WEIGHT display will show flashing XX and then
changes to XX.XXX lb. Once this occurs the calibration process is complete.
-
***Note: Make sure nothing is loaded on the platform before you are asked to place the
calibration weight. Once calibration is complete you will need to reboot the scale and place the
sample test weight on the platter to assure the calibration was completed correctly.
5.3 Scale 3 Calibration
- First make a connection to SCALE 3 through DB9
SCALE 3 DB9
1 S-
2 S+
3/6 V+
4 V-
5 GND
7-9 NC
Page 14 of 16
Load Cell DB9
1 S-
2 S+
3/6 V+
4 V-
5 GND
7-9 NC
AL-manual-EC200-En-V2