Table 4.4.1 Suggested A2A-D2 Settings
Output Signal
Excitation
0-10V ( 0±10V)
4-20mA
0-5V ( 0±5V)
Note: When applicable, it is recommended to choose 12V excitation and configure Gain switches either 2, 3 or 4
Table 4.4.2 Gain Switch Settings
4-bit DIP Switch
GAIN
Note: If the Gain setting is not achieved during calibration, change the Gain switch settings as shown in Table 4.4.2
and recalibrate.
5.

Operations

A2A-D2 calibration has steps of Gain selection namely Zero calibration and
Span calibration.
5.1 Gain Selection
1. Connect the load cell. Set the slide switch EXC and select load cell
excitation voltage
2. Based on the excitation voltage chosen and the load cell's sensitivity
( refer to load cell's documentation and refer to Table 4.4.1 ), set
the recommended Gain Switch to select Gain range
4 |
ANYLOAD A2A-D2 Amplifier Product Manual (V1810)
Sensitivity Range
( mV/V)
1.30-4.20
1.35-2.60
2.68-3.90
0.54-1.89
1.51-2.91
2.96-4.31
0.61-2.02
0.66-1.26
1.30-1.89
1.67-2.25
1.56-3.02
3.09-4.54
0.64-2.13
0.68-1.32
1.35-1.97
1.74-2.36
No. of DIP Switch
1
2
3
4
Voltage (V)
5
12
5
12
5
12
Gain Range
Gain Switch
Number
4
1
2
4
1
2
4
1
2
3
1
2
4
1
2
3
217-753
153-303
103-153
86-118