Emergency stop
switch
PW 811 / PW 814 / PW 818
 For washing machines without a separate emergency stop
switch, there must be a central emergency shut-off at the
installation site.
The emergency stop switch puts the machine into a safe state in the
event of danger or in order to prevent danger. The emergency stop
switch must be positioned in such a way that the user can activate it
immediately in an emergency.
If it is necessary to install a residual current device (RCD) in
accordance with local regulations, a residual current device type B
(sensitive to universal current) must be used.
An existing type A residual current device (RCD) must be exchanged
for a type B RCD.
If necessary, equipotential bonding with a good contact
connection must be guaranteed in compliance with all applicable
local and national installation specifications.
Equipotential bonding must have an earth current rating > 10 mA.
Accessories for equipotential bonding are not supplied and need to
be ordered separately.