DIAVITE AG DIAVITE Compact VH 매뉴얼 - 페이지 7

{카테고리_이름} DIAVITE AG DIAVITE Compact VH에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. DIAVITE AG DIAVITE Compact VH 14 페이지.

DIAVITE AG DIAVITE Compact VH 매뉴얼
1.7

Calibration

The DIAVITE COMPACT has an automatic calibration, which is at disposal in the menu 3.1 -
'configuration - calibration' (page 10). In order to avoid faulty measurements, it is absolutely
indispensable to check the perfect function of the instrument before starting the first daily
application, as well as in regular periods of time. This can be effected the easiest way with a
roughness specimen 3.00 µm Ra, alternatively with another reference specimen, the exact
value of which is known.
When changing a tracer, the instrument has to be recalibrated also. The deviations between the
various tracers should be rather small, normally, but could cause anyhow faulty measuring
results. The recalibration of the instrument runs automatically. Put the reference specimen
onto the supporting shoe of the traverse unit or effect this procedure employing a measuring
support. The settings must be correct.
1.
Press any button, after some seconds, the instrument is ready for operation
2.
Press the button
3.
Press the button
4.
Press the button
5.
Press the button
For calibration, the cutoff 0.8 mm will be set automatically!
6.
Press the button
7.
Press the button
8.
It takes approximately one minute to effect the calibration (4 measurements)
9.
At the end of the calibration procedure, the current measuring value Ra is indicated. In course,
the instrument is in the 'main menu' again
10.
Press the button
1.8

Keyboard

The DIAVITE COMPACT is a surface roughness meter easy to operate and for the safe operation of
which only few introduction is necessary. The instrument can be set according to the operator's needs
using the menu functions. The keys are provided with pictographs for the following functions:
Measuring mode
Start of measuring operation
Selection of length of traverse
Menu for instrument configuration
Interrogation of measuring value
The emergency stop after release of the measurement can be effected with any key
DIAVITE COMPACT
, you are in the mode 'Menu: Tracer'
until the indication 'Menu: Config.' appears
once, in the display appears 'Config.: Calibration'
, the indication '3.00' appears
or
, if you do not employ the original roughness standard
once, for starting the calibration
, you are again in the measuring mode
Menu mode
Confirmation
Menu selection previous menu option
Switch-on/off menu, Escape
Menu selection next menu option
6