the clamp on side of the microscope.
Position back plate so that vertical adjustment screws are midway in slots
Use the 5/64" Allen Wrench to tighten the Base plate clamp at a point where the vertical adjustment
screws are midway in the slots
After securing the back plate check for proper drive alignment be rotating the right fine focus knob.
The drive is usually self aligning and when properly aligned no drag is felt, other then that caused by
the gears, while rotating the fine focus knob. When the dive is in a position of proper alignment use
the 5/32 Allen wrench to tighten the horizontal, and vertical adjustment screws at the position where
minimum drag is felt.
Recheck for minimum drag on the fine focus shaft by turning the right fine focus knob. If there is any
drag felt at any particular point while rotating the right fine focus knob repeat the alignment
procedure by loosening the horizontal, and vertical adjustment screws and repositioning the drive.
The drive is usually self aligning , and if realignment is necessary loosen the horizontal, and z vertical
adjustment screws and move the drive so that is slides into its natural position and then tighten the
Applied Scientific Instrumentation - https://asiimaging.com/docs/