Olympus BX2 SERIES 유지 관리 매뉴얼 - 페이지 23

{카테고리_이름} Olympus BX2 SERIES에 대한 유지 관리 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BX2 SERIES 29 페이지. Olympus microscope bx2
Olympus BX2 SERIES에 대해서도 마찬가지입니다: 정보 (6 페이지)

Olympus BX2 SERIES 유지 관리 매뉴얼
BX2 MAINTENANCE MANUAL
(5) Adjustment procedure (BX41)
Connect to the
connector CN21 on
power board.
Light preset
switch
Connect to the
rheostat ass'y (A011).
(6) Adjustment procedure(BX45)
Connect to the connector
CN21 on power board.
Connect to the
rheostat ass'y (A011).
REPAIR PROCEDURE
VR203
VR202
B
Control board
Light preset switch
VR203
VR202
Control
board
1) Turn on the power, and then turn the
knob "B" (P.18) to the maximum level.
2) While turning the trimmer (VR203"A") on
the control board, adjust the lamp output
voltage to 5.9V using the digital multimeter.
A
Standard
3) Turn on the light preset switch.
4) Set the light preset voltage to 4V by
turning the the trimmer (VR202"B").
Standard
5) For prevention of turning, apply
solvent-based adhesive(OT1026) on
the one spot of trimmer(VR203).
1) Turn on the power, and then turn the
knob "B" (P.18) to the maximum level.
2) While turning the trimmer (VR203"A") on
the control board, adjust the lamp output
voltage to 5.9V using the digital multimeter.
A
Standard
3) Turn on the light preset switch.
4) Set the light preset voltage to 4V by turning
the trimmer (VR202"B").
B
Standard
5) For prevention of turning, apply
solvent-based adhesive(OT1026) on the
spot of trimmer (VR203).
- 20 -
DC5.9V+/-0.1V
DC4V+/-0.1V
DC5.9V+/-0.1V
DC4V+/-0.1V