Olympus BX2 SERIES 유지 관리 매뉴얼 - 페이지 21

{카테고리_이름} Olympus BX2 SERIES에 대한 유지 관리 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus BX2 SERIES 29 페이지. Olympus microscope bx2
Olympus BX2 SERIES에 대해서도 마찬가지입니다: 정보 (6 페이지)

Olympus BX2 SERIES 유지 관리 매뉴얼
BX2 MAINTENANCE MANUAL

4. Electrical adjustment

The following shows the voltage adjustment when the rheostat ass'y is replaced.
(1) Replacement of rheostat ass'y
Rheostat ass'y
(A011/A012)
(2) Setup for adjustment (BX51/52)
VR52
REPAIR PROCEDURE
E
D
C
Knob
(A010)
Connect the lamp housing.
(U-LH100-3/U-LH100)
Power board
CN51
Lamp voltage
output terminal
12.2
Digital multimeter(RMS)
1) Remove the bottom cover "A" of
microscope frame.
2) Loosen the screw of knob "B" and
disassemble the parts in order of "B" to
to "E".
Note: Remove the rheostat ass'y
3) Disconnect the connector of rheostat
ass'y.
4) Reassemble them in reverse order.
B
1) Remove the back panel of microscope
frame.
2) Remove the control board . (Refer to the
next page.)
3) Prepare for adjustment as shown in the
left figure.
- 18 -
from inside of base.
Screw: CUK3X6SA
Screw: CUTB3X6SA
8 pcs.
4 pcs.